TEM interactive laboratory simulators
Experiment with relativistic momentum, reciprocal space, and objective aperture optics
SAED electron diffraction simulator
Selected-Area Electron Diffraction (SAED) & Reciprocal Space Geometry
CAMERA CONSTANT: 1.50 mm·nm
Diffraction controlsTEM column
Camera Length (L)
600 mm
200 mm700 mm1200 mm
Increasing camera length optically magnifies the diffraction pattern.
Voltage (HT)
200 kV
50 kV175 kV300 kV
Higher voltage decreases electron wavelength λ, moving diffraction spots closer to the center.
TEM phosphor viewing screen / diffraction patternDetector plane
ZONE AXIS [001]
λ (Wavelength)
2.51 pm
Camera length (l)
600 mm
Camera constant (λl)
1.505 mm·nm
WHAT CHANGED? (Williams & Carter Reciprocal Relationships)
- →Camera Constant equation:
R · d = λ · L. The physical radius R of any diffraction spot on the detector is inversely proportional to lattice spacing d and directly proportional to camera length L. - →Increasing Camera Length (600 mm): Diffraction spots move farther from the direct beam (000), increasing angular dispersion.
- →Increasing Voltage (200 kV): Relativistic momentum increases, reducing electron wavelength λ. Consequently, diffraction angles 2θ_B shrink, drawing all spots inward.